TR-TCSP04-1A Single Channel Syringe Pump


COMPANY : Tryte Technology (H.K.) Limited
ADD : UNIT 04, 7/F, BRIGHT WAY TOWER, NO. 33 MONG KOK ROAD, KOWLOON, HONGKONG.
TEL:00852-2389298
FAX: 00852-35902333
CELL:0086-15367067787(Robin)
SITE : www.trytetc.com
SKYPE : live:myformat666
EMAIL : sale@trytetc.com
more>>>

  • Name: TR-TCSP04-1A Single Channel Syringe Pump
  • Model : TR-TCSP04-1A
  • Erp code: 9000000388
  • | Share

TR-TCSP04-1A.png

Instruction:

This laboratorial syringe pump is a four channels syringe pump which only has infusion mode. The acceptable syringe specification is from 10 μL to 10 mL. Suitable for high accuracy and small flow rate liquid transferring.

 

Specification:

Syringe pump

TR-TCSP04-1A

No. of Syringes

1

Advance per Mircostep

1/16step: 0.156 μm

Infusion Volumn per Microstep

0.026 μL (10 mL BD Syringe)

Syringe Size

10 μL - 10 mL

Max. Linear Rate

130 mm/min

Min. Linear Rate

5μm/min

Flow Rate

0.831 nL/min -  21.675 ml/min

Working Mode

Infusion

Linear Force

> 9 kgf

Max. Step Rate

13867  (1/16step)/ sec

Min. Step Rate

16  (1/16 step) /30 sec

Accuracy

≤ 0.5% error in the condition of > 30% of max. infusion distance

Setting Mode

Rotary coded switch and membrane keypad

Display

128 x 64 graphic LCD

Power

AC 100 - 240 V

Operating Condition

Temperature  5 °C - 40 °C       Relative humidity < 80 %

Dimensions

280 × 250 × 140 mm

Weights

4.5 kg

 

Functions and features:

- Syringe Selection: The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly 
- Easy to operate: Combining big screen LCD display with rotary coded switch and membrane keypad makes the operation simple and prompt. 
- Working Mode: Infusion 
- Memory Function: 
1. The parameters are saved in EEPROM. The parameters don't need to be reset when power returns after an interruption 
2. In flow rate mode, the pump remains running or stop according to the setting parameters when power returns after an interruption 
- Protection Function: The pump will stall and give an alarm when the drive structure of the pump is blocked 
- Communication Function: Realize computer control through RS485 communication interface 
- External Control Function: Input/output control 
- Calibration Function: Acquire accurate volume through calibration 
- Syringe Protection: Adjust syringe rest to prevent syringe from damaging

 

Parameters:

TypeProduct CodeSyringeSyring ID (mm)Flow Rate (μl/min-ml/min)SP-LSP04-1A050345110μl0.500.001-0.025525μl0.800.0025-0.065350μl1.100.0048-0.1235100μl1.600.0101-0.2614250μl2.300.0208-0.5401500μl3.250.0415-1.07841ml4.720.0875-2.27472ml9.000.3181-8.27025ml13.100.6739-17.52210ml16.601.0821-28.135


Specification:

Syringe pump

TR-TCSP04-1A

No. of Syringes

1

Advance per Mircostep

1/16step: 0.156 μm

Infusion Volumn per Microstep

0.026 μL (10 mL BD Syringe)

Syringe Size

10 μL - 10 mL

Max. Linear Rate

130 mm/min

Min. Linear Rate

5μm/min

Flow Rate

0.831 nL/min -  21.675 ml/min

Working Mode

Infusion

Linear Force

> 9 kgf

Max. Step Rate

13867  (1/16step)/ sec

Min. Step Rate

16  (1/16 step) /30 sec

Accuracy

≤ 0.5% error in the condition of > 30% of max. infusion distance

Setting Mode

Rotary coded switch and membrane keypad

Display

128 x 64 graphic LCD

Power

AC 100 - 240 V

Operating Condition

Temperature  5 °C - 40 °C       Relative humidity < 80 %

Dimensions

280 × 250 × 140 mm

Weights

4.5 kg

 