TR-TCLC-02 High Performance Liquid Chromatography(HPLC)


COMPANY : Tryte Technology (H.K.) Limited
ADD : UNIT 04, 7/F, BRIGHT WAY TOWER, NO. 33 MONG KOK ROAD, KOWLOON, HONGKONG.
TEL:00852-2389298
FAX: 00852-35902333
CELL:0086-15367067787(Robin)
SITE : www.trytetc.com
SKYPE : live:myformat666
EMAIL : sale@trytetc.com
more>>>

  • Name: TR-TCLC-02 High Performance Liquid Chromatography(HPLC)
  • Model : TR-TCLC-02
  • Erp code: 9000000347
  • | Share

TR-TCLC-02 .jpg

Application:

High-performance liquid chromatography (or high-pressure liquid chromatography, HPLC) is a chromatographic technique that can separate a mixture of compounds, and is used in biochemistry and analytical chemistry to identify, quantify and purify the individual components of the mixture.

sampling split system and ultraviolet visible detector constitute this machine.


Features:

TR-TCLC-02 HPLC adopts microcomputer high pressure infusion pump with two complementary pumps and self-examine system. The materials of the pump are stainless steel or Polymer Materials.

Ultraviolet visible detector uses Grating system. Specially --made Micro-- flowing cell can set time-varying wavelength analyze.   

The terminal device is five inches colored LCD which can observe the results of the observed testaments and spectrogram; it can also function as the inputting window for the operation parameters.

The international used standard six-way valve  is  chose  as  the  sample injector; the materials of the connect cannula is stainless steel or PEEK cannula;  separator column use standard C18 column; The instrument has a column thermostat, the case can install two 70mm-250mm column tubules at the same time. The canuals,  joint washers and Spindle Adapters used in the analyze system are all meet the international requirements. It can share the out porting signals with the import instruments. It has data processing machine interface (1AU/V) and recorder interface (10mV/FS) and equipped with RS232 output interface which can cooperate with chromatographic working station to achieve two-way alternate communication control. 
 

Performance Indicators:

Chromatogram column thermostat:

Rang of temperature control: with a room temperature from 10 to 89(increment 1°C)

Temperature stability error should below 0.5 °C

High pressure pump:

Flow rate range: 0.1-8ml/min

Flow rate stability error ≤2%

Flow rate precision≤5%

The maximum output pressure:

metal system: 40MPa

non-metallic system: 25MPa

Ultraviolet visible detector:

The range of the wavelength: 190nm-700nm

The accuracy of the wavelength: ±2nm

Absorbance relative error: ±10%

Baseline noise≤1×10-4AU (254nm)

Baseline drift≤1×10-3AU/h (254nm)


Performance Indicators:

Chromatogram column thermostat:

Rang of temperature control: with a room temperature from 10 to 89(increment 1°C)

Temperature stability error should below 0.5 °C

High pressure pump:

Flow rate range: 0.1-8ml/min

Flow rate stability error ≤2%

Flow rate precision≤5%

The maximum output pressure:

metal system: 40MPa

non-metallic system: 25MPa

Ultraviolet visible detector:

The range of the wavelength: 190nm-700nm

The accuracy of the wavelength: ±2nm

Absorbance relative error: ±10%

Baseline noise≤1×10-4AU (254nm)

Baseline drift≤1×10-3AU/h (254nm)



0