Distillate Fuel Cold Filter Plugging Point Filter


COMPANY : Tryte Technology (H.K.) Limited
ADD : UNIT 04, 7/F, BRIGHT WAY TOWER, NO. 33 MONG KOK ROAD, KOWLOON, HONGKONG.
TEL:00852-2389298
FAX: 00852-35902333
CELL:0086-15367067787(Robin)
SITE : www.trytetc.com
SKYPE : live:myformat666
EMAIL : sale@trytetc.com
more>>>

  • Name: Distillate Fuel Cold Filter Plugging Point Filter
  • Model : TR-TC-1
  • Erp code: 8000000035
  • | Share

The instrument is designed and made as per the Industry Standard of People’s Republic of China SH/T 0248 “Test Methods for Cold Filter Plugging Point of Distillate Fuels”. It is suitable to be used to determine cold filter plugging point of diesel engine fuels and crude diesel oils, including fuels containing flow improver, as per SH/T0248, together with cooling bath and other device.
 

Main technical characteristics:
1. It adopts integrated structure. It is novel, compact and beautiful.
2. It adopts touch button and digital display technologies, so it is easy to preset parameters. The timing time can be shown clearly and the instrument will stop timing when the timing time is over.
3. It is equipped with complete accessories. These accessories are equipped as per the requirements of SH/T0248 “Test Methods for Cold Filter Plugging Point of Distillate Fuels”.
4. The instrument can be used independently for cold filter plugging point test together with proper cooling bath or used for cold filter plugging point test as an accessory of relative low temperature tester.
 

Main technical specification and parameters:
1. Power supply: AC 220V±10%, 50 Hz;
2. Pumping pressure: U type differential pressure meter 1961 Pa (200 mm H2O);
3. Sample cup:
(1) Round glass cup with flat bottom;
(2) Inner diameter: 31.0~32.0㎜;
(3) Wall thickness: 1.0~1.5 mm;
(4) Height: 115~125㎜;
(5) There is a mark line at the 45 ml of the cup.
4. Filter: It is made of copper. Filter No. is 004 (363 mesh);
5. Working environment:
(1) Ambient temperature: -10~+35℃;
(2) Relative humidity: ≤85%;
6. Total power consumption: not more than 150 W;


PT-LC-1.jpg

Main technical specification and parameters

1. Power supply: AC 220V±10%, 50 Hz;
2. Pumping pressure: U type differential pressure meter 1961 Pa (200 mm H2O);
3. Sample cup:
(1) Round glass cup with flat bottom;
(2) Inner diameter: 31.0~32.0㎜;
(3) Wall thickness: 1.0~1.5 mm;
(4) Height: 115~125㎜;
(5) There is a mark line at the 45 ml of the cup.
4. Filter: It is made of copper. Filter No. is 004 (363 mesh);
5. Working environment:
(1) Ambient temperature: -10~+35℃;
(2) Relative humidity: ≤85%;
6. Total power consumption: not more than 150 W;


0