TR-TC-755B CRT UV-VIS Spectrophotometer 200-1100nm, 2nm, LCD


COMPANY : Tryte Technology (H.K.) Limited
ADD : UNIT 04, 7/F, BRIGHT WAY TOWER, NO. 33 MONG KOK ROAD, KOWLOON, HONGKONG.
TEL:00852-2389298
FAX: 00852-35902333
CELL:0086-15367067787(Robin)
SITE : www.trytetc.com
SKYPE : live:myformat666
EMAIL : sale@trytetc.com
more>>>

  • Name: TR-TC-755B CRT UV-VIS Spectrophotometer 200-1100nm, 2nm, LCD
  • Model : TR-TC-755B
  • Erp code: 9000000156
  • | Share



TR-TC-755B1 拷贝.jpg


Features:

l   This instrument adopts 128 * 64-bit dot matrix LCD, can directly show the standard curve and test data, the main engine can save test data, and choose printer.

l   USB data output interface, optional 3.0 professional software for on-line operation;

l   Able to establish standard curve directly, and use it for relevant tests. May test and save 200 groups of data continuously, and save 100 specification curves, user can name directly, and transfer according to naming convenience.

l   Optical system with unique design, high-performance 1200 bar/mm grating and receiver ensures the instrument with excellent performance index.

l   Automatic wavelength calibration, automatic wavelength setting, automatic switching illuminant, automatic controlling deuterium lamp and tungsten lamp's switch, real-time monitoring lamp light time.

l   Wide sample room can hold 5-100mm various specifications of cuvettes.

l   Film buttons, convenient and simple operation.

l   Use 3.0 professional analysis software, can realize the standard curve, dynamic test and other functions.

 


755-1.jpg

755-2.jpg


Technical Parameters:

Wavelength range:

190- 1100 nm

Wavelength accuracy:

+/- 1nm

Spectral bandwidth:

2 nm

Wavelength repeatability:

0.2 nm

Stray light:

<= 0.1%T @220 nm, 360 nm

Photometric accuracy:                   

0.5%T

Photometric repeatability:

0.2%T

Photometric range

0~200%T, -0.3-1~3.000A, 0~9999C

Transmittance accuracy

+/-0.002A(0~0,5A); +/-0.003A(0.5~1A);+/-0.3%T

Transmittance repeatability

<=0.001A(0~0.5A); <=0.002(0.5~1A); <=0.15%T(0~100%T)

Drift

<=0.001A/h@500nm

Baseline linear

+/-0.002A/h@500nm

Stability:

+/- 0.002 A/h @500 nm

Work Pattern:

T, A, C

Display area:

0- 200%T, -0.3-3A

Display modes:

128*64 LCD

Zero setting:

Automatic

Weight:

12 kg

 755-3.png

Accessories:

Model

TR-TC-755B 

Host

Standard accessories

Power line

Standard accessories

Instruction manual

Standard accessories

Certification

Standard accessories

Packing list

Standard accessories

Warranty card

Standard accessories

Glass cuvette1CM

4Pcs

Quartz cuvette1CM

2Pcs

5CMComparator Block

Optional

Printer

Optional

Software

Optional

2cm,3cm,5cm cuvette

Optional

Tungsten lamp12V20W

Optional

Deuterium lamp

Optional


Technical Parameters:

Wavelength range:

200- 1100 nm

Wavelength accuracy:

+/- 1nm

Spectral bandwidth:

2 nm

Wavelength repeatability:

0.5 nm

Stray light:

<= 0.1%T @220 nm, 360 nm

Photometric accuracy:                   

0.5%T

Photometric repeatability:

0.2%T

Photometric range

0~200%T, -0.3-1~3.000A, 0~9999C

Transmittance accuracy

+/-0.002A(0~0,5A); +/-0.003A(0.5~1A);+/-0.3%T

Transmittance repeatability

<=0.001A(0~0.5A); <=0.002(0.5~1A); <=0.15%T(0~100%T)

Drift

<=0.001A/h@500nm

Baseline linear

+/-0.002A/h@500nm

Stability:

+/- 0.002 A/h @500 nm

Work Pattern:

T, A, C

Display area:

0- 200%T, -0.3-3A

Display modes:

128*64 LCD

Zero setting:

Automatic

Weight:

12 kg

 


0