TR-AG6700 Automatic Proximate Analyzer


COMPANY : Tryte Technology (H.K.) Limited
ADD : UNIT 04, 7/F, BRIGHT WAY TOWER, NO. 33 MONG KOK ROAD, KOWLOON, HONGKONG.
TEL:00852-2389298
FAX: 00852-35902333
CELL:0086-15367067787(Robin)
SITE : www.trytetc.com
SKYPE : live:myformat666
EMAIL : sale@trytetc.com
more>>>

Your present location: Home > Product> Coal&Steel Testing> Proximate Analyzer>
  • Name: TR-AG6700 Automatic Proximate Analyzer
  • Model : TR-AG6700
  • Erp code: 2000000034
  • | Share

TR-AG6700 Automatic Proximate Analyzer(also called thermogravimetric analyzer, TGA)is used to determine the moisture, ash, volatile matter in coal, coke, biomass, limestone and iron ore,food and feedstuffs by thermogravimetric method. It can also determine residual ash and slag combustibles in accordance with ASTM, which is widely applied in power plants, coal mines, metallurgy, food industry,chemical industry, commercial inspection, scientific research, education, etc. It could work both simultaneously and independently.

TR-AG6700.jpg


Standard Configuration       
1.Computer  2.Printer  3.Main analyzer: Part I and Part II  4.Air compressor  5.Volatile matter Crucibles
 6.Ash Crucible 7.Weighing Rod Base  8.Weighing Rod  9.Three-Core Power Socket and Plug/20A  
10.Crucible Lid for Volatile Matter 11.Glass Spoon(small size)  12.Standard Coal Sample (GBW)  13.Tool kit

 

Features
High Stability and Reliability
1. Dual furnace which could work both simultaneously and independently. Unique isolation board to ensure accurate result of volatile matter. 
2. Patented technology of heat-resistant to ensure stability and reliability.
3. Built-in balance ensures unbeatable accuracy, comply with thermogravimetric method.
4. Unique design to ensure the stability of heating efficiency and temperature keeping. 
5. Using blank crucible to calibrate the influence of the thermal buoyancy. 

 

Convenient Operation
1.The operator is limited to just adding sample to a crucible.
2. Reference method and customized method available. 
3. Capable to load sample continuously and singly.
4. Real time display of balance reading, which makes it convenient to control the precision of sample weight.

 

Cost Saving
Available for analysis without oxygen and nitrogen by configured air pump, except for lignite.

 

Good Safety
Automatically heating off when open the ash furnace cover.

 

TR-AG6700-1.jpg

Intelligent Software Systems

14597768305097.jpg


MODELTR-AG6700 Automatic Proximate Analyzer
Conforms to MethodASTM D7582, ISO17246 and GB/T212
Max. Sample Loading19
FurnaceDual furnace
Analysis Time≤120mins for 19 samples
Sample Mass0.8-1.2g recommended  / up to 5g
Temp. RangeUp to 1050°C
Temp. Control Precision ± 3 °C
Precision of Balance0.1mg
Power SupplySingle phase, AC220±10%, 50/60Hz, Part I: ≤ 4 kW
Part II: ≤ 5 kW
Net WeightPart I: 80kg    Part II: 50kg
Dimensions (L×W×H)Part I: 550mm×580mm×890mmPart II: 550mm×580mm×530mm