Application:
It is applied to determine the fluorine and chlorine content in coal or other combustible material.
Specifications:
Test range:Fluorine:10ppm〜2000ppm Chlorine:30ppm〜3000ppm
Test time:Including calibration of electrode parameters, sample decomposition and collection, titration
Calibration of electrode parameters:Available to calibrate when decomposing the first batch of samples and not calculated to total analysis time. Not necessary to calibrate each day, calibration be be proceeded after long-time test. If electrode is not used for long time, calibration is also needed.
Decomposition:30min approximately
Titration:5min approximately for fluorine;10min approximately for chlorine
Sample weight:0.5g
Temperature control precision:1100±5℃
Max power:4kW
Repeatability:Conform to GB standard
Uncertainty:Within uncertain region of standard sample
Specifications:
Test range:Fluorine:10ppm〜2000ppm Chlorine:30ppm〜3000ppm
Test time:Including calibration of electrode parameters, sample decomposition and collection, titration
Calibration of electrode parameters:Available to calibrate when decomposing the first batch of samples and not calculated to total analysis time. Not necessary to calibrate each day, calibration be be proceeded after long-time test. If electrode is not used for long time, calibration is also needed.
Decomposition:30min approximately
Titration:5min approximately for fluorine;10min approximately for chlorine
Sample weight:0.5g
Temperature control precision:1100±5℃
Max power:4kW
Repeatability:Conform to GB standard
Uncertainty:Within uncertain region of standard sample