TR-IRS3600 Automatic Infrared Sulfur Analyzer


COMPANY : Tryte Technology (H.K.) Limited
ADD : UNIT 04, 7/F, BRIGHT WAY TOWER, NO. 33 MONG KOK ROAD, KOWLOON, HONGKONG.
TEL:00852-2389298
FAX: 00852-35902333
CELL:0086-15367067787(Robin)
SITE : www.trytetc.com
SKYPE : live:myformat666
EMAIL : sale@trytetc.com
more>>>

Your present location: Home > Product> Coal&Steel Testing> Sulfur Analyzer>
  • Name: TR-IRS3600 Automatic Infrared Sulfur Analyzer
  • Model : TR-IRS3000
  • Erp code: 2000000017
  • | Share

The TR-IRS3600 Automatic Infrared Sulfur Analyzer is used to determine the total sulfur content by infrared absorption ,which is widely applied in power plants ,coal mines ,metallurgy ,chemical industry ,commercial inspection ,scientific research ,etc . 


TR-IRS3600.jpg


Standard Configuration 
1.Computer  2.Printer  3.Main analyzer  4.A/C adapter  5.Anhydrone  6.Silica wool
7.Crucibles   8.Outer combustion tube  9.Inner combustion tube  10.Boat stop  11.O-ring kit   12.Tool kit
 
Features
High Automation
1.Unique dual oxygen lance to increase the combustion efficiency
2.Maximum throughput with auto-loader ,available for 60 samples per batch and easy to add or reduce sample quantity during analysis.
3.Patented sample delivering system realizes smooth work flow of pulling and discharging samples.
4.The sample mass can be automaticlly sent to the computer by balance connection.
5.Available for unattended operation by intelligent sensor. 
6.Auto self-diagnosis and language alert.
 
Accuracy and Stability 
1. Top quality ultra-low drift infrared cell to ensure stability of the testing results. 
2. Reliable single Si-C spiral tube heating components. 
3. Unique gas tightness structure to avoid the effect of SO2 in air.
 
Environmental Protection
Exhaust emissions ventilation system to avoid air pollution of laboratory.
 

TR-IRS3000-automatic-infrared-sulfur-analyzer-software-system.jpg

 


MODELTR-IRS3000 Automatic Infrared Sulfur Analyzer
Conforms to MethodASTM D4239, ISO19579 and GB/T25214
Max. Sample loadingUp to 60 samples per batch automatically
Analysis MethodInfrared absorption 
Analysis Resolution0.001%
Sulfur Range0.01%-30% Customized range available
Analysis Time per Sample≤120s
Analysis Temp.1350°c
Temp.Control Precision±1°c
Sample Mass200mg-400mg for coal and coke (300mg is recommended)
Power SupplySingle phases, AC220±10%, 50/60Hz, ≤4kW 
Net Weight70kg
Dimensions(L×W×H)750mm×650mm×700mm

 


0