Overall Performance:
Equipment type:Fully automatic discrete, STAT priority
Analysis rate:Constant speed 300T/H(without ISE)
Test principle:Colorimetry, turbidimetry
Analysis method:End-point, 2 kinetics, fixed-time, etc. support single/double wavelength, linear and non-linear calibration
Sample System:
Sample Tray: 60 sample positions, random access of all samples
Sample Volume: 2-50ul, 0.1ul increment
Sample probe: Liquid level detection, the remaining detection, collision protection etc.
Sample Containers: Sample cups, Micro tube, Original blood-collection tube, etc.
Reagent System:
Reagent Tray:45 reagent positions
Reagent Volume: R1:25-300ul, R2:10-150ul, 0.1ul increment
Reagent Probe: Liquid level detection, the remaining detection, collision protection etc.
Reagent Refrigeration:4-15℃, independent power supply, 24-hour non-stop refrigeration
Reaction System:
l Reaction Tray Rotating tray, 120 cuvettes
l Cuvette Reusable high quality cuvette, Optical diameter 6mm
l Reaction Volume 180-500ul
l Incubation system Bi-directional liquid-air device
l Temperature 37℃±0.1℃ with real-time display
Optical System:
l Light source 12V35W halogen lamps, over 2000 hours
l Photoelectron road after spectrophotometry
l Wavelength 340,405,450,510,546,578,630,700nm
l Absorbance range 0-3.500Abs
l Resolution 0.0001Abs
Operating System:
l PC operating system Windows XP or above
l Analysis control software English version graphical operating software
l Report printing Report formats support the user-defined mode, QC and state information etc.
l System connection RS-232
Working Conditions:
n Power supply: AC 220V±22V, 50Hz±1Hz, power 1.5KVA
n Ambient Temperature: 15-30 ℃
n Humidity: 35-80%
n Water consumption: 7L/Hour
n Dimension: Bench top: 750mm*520mm*600mm
Overall Performance:
Equipment type:Fully automatic discrete, STAT priority
Analysis rate:Constant speed 300T/H(without ISE)
Test principle:Colorimetry, turbidimetry
Analysis method:End-point, 2 kinetics, fixed-time, etc. support single/double wavelength, linear and non-linear calibration
Sample System:
Sample Tray: 60 sample positions, random access of all samples
Sample Volume: 2-50ul, 0.1ul increment
Sample probe: Liquid level detection, the remaining detection, collision protection etc.
Sample Containers: Sample cups, Micro tube, Original blood-collection tube, etc.
Reagent System:
Reagent Tray:45 reagent positions
Reagent Volume: R1:25-300ul, R2:10-150ul, 0.1ul increment
Reagent Probe: Liquid level detection, the remaining detection, collision protection etc.
Reagent Refrigeration:4-15℃, independent power supply, 24-hour non-stop refrigeration
Reaction System:
l Reaction Tray Rotating tray, 120 cuvettes
l Cuvette Reusable high quality cuvette, Optical diameter 6mm
l Reaction Volume 180-500ul
l Incubation system Bi-directional liquid-air device
l Temperature 37℃±0.1℃ with real-time display
Optical System:
l Light source 12V35W halogen lamps, over 2000 hours
l Photoelectron road after spectrophotometry
l Wavelength 340,405,450,510,546,578,630,700nm
l Absorbance range 0-3.500Abs
l Resolution 0.0001Abs
Operating System:
l PC operating system Windows XP or above
l Analysis control software English version graphical operating software
l Report printing Report formats support the user-defined mode, QC and state information etc.
l System connection RS-232
Working Conditions:
n Power supply: AC 220V±22V, 50Hz±1Hz, power 1.5KVA
n Ambient Temperature: 15-30 ℃
n Humidity: 35-80%
n Water consumption: 7L/Hour
n Dimension: Bench top: 750mm*520mm*600mm