Product Description:
Adopting microcomputer control digital broadband electro-liquid servo valve to drive precise hydraulic cylinder, this machine can conduct tensile, compression, bending-resistance tests through automatic control of test force, displacement, deformation and other models. It is in line with the national standard GB/T228-2002 "Room-temperature Tensile Test Methods of Metal Materials” and other standards.
The host of testing uses hydraulic cylinders mounted; with the device to adjust the test space and transmission equipment installed at the bottom, stable transmission; open structure of the upper and lower jaw to facilitate clamping samples and of good stability.
Independent hydraulic servo loading system and high-precision broadband servo valve ensure high precision, high efficiency, low noise and fast response; Independent hydraulic clamping system ensures smooth operation with low noise and solid clamping during the testing process.
Lenovo computers are applied to complete test parameters setting, test process controlling, data acquisition, processing, analysis, storage and display (test data includes: upper and lower yield point, tensile strength, tensile strength, elastic modulus volume and point elongation, etc.) during the testing process. It is of high precision and accuracy in controlling and measuring.
Technical parameters of controlling:
1. using high-precision magnifier and 24-bit high-precision AD converter, the effective internal code: ± 120000; testing force resolution of 1 / 120000.
2. using PID control to complete constant speed control of load, speed range: 0.2-5% FS / S has the function of load protection
3. using PID control to complete constant speed control of displacement, speed range: 0.6-60mm / min has the function of load protection
4. using PID control to complete constant speed control of strain, speed range: 0.00025/S-0.0025/S has the function of load protection
Technical Parameters:
Model Items | TR-HUT-10 |
Max. load | 600kN |
Measuring range | 2%-100%Fs(or 0.6-100%) |
Relative error of reading | ±1% better than reading |
Testing force resolution | 1/120000 |
Max. distance between tensile jaw (including the piston stroke) | 600mm |
Max. clamping width of flat specimen | 70mm |
Clamping thickness of flat specimen | 0-30mm |
Clamp diameter of round specimen | Φ13-40mm |
Pump motor speed | 1.5kW |
Motor power of crosshead up and down | 0.5kW |