Summary:
The instrument is designed and made as per national standard of People’s Republic of China SH/T0175 Standard test method for oxidation stability of distillate fuel oil (accelerated method) and ASTM D2274 Standard Test Method for Oxidation Stability of Distillate Fuel Oil (Accelerated Method).It is suitable to determine the oxidation stability of distillate fuel oils with accelerated method.
The instrument adopts metal bath structure. No need to add water during determination. Clean and eco-friendly. Easy to operate. It is the preferred automated instrument to determine the oxidation stability of distillate fuel oils for the units of petroleum mining, producing and using and other relevant colleges and scientific research institutions.
Main technical features:
1. Floor type and unit-body structure. Stainless steel working table. Easy to use.
2. Metal bath. No need to add water during determination. No pollution and energy saving.
3. Six test baths and six flow meter. It can determine six samples at a time. Working efficiency is high.
4. The heating rate can be adjusted continuously.
Main technical specifications:
1. Power supply: AC 220 V±10%, 50 Hz
2. Maximum power consumption: 2400W
3. Temperature control mode: Automatically controlled by digital temperature controller
4. Temperature control range: Ambient to 200℃
5. Temperature control accuracy: ±0.2℃
6. Temperature measuring component: Thermal resistance
7. Sample quantity:It can determine 6 samples at a time
Main technical specifications
1. Power supply: AC 220 V±10%, 50 Hz
2. Maximum power consumption: 2400W
3. Temperature control mode: Automatically controlled by digital temperature controller
4. Temperature control range: Ambient to 200℃
5. Temperature control accuracy: ±0.2℃
6. Temperature measuring component: Thermal resistance
7. Sample quantity:It can determine 6 samples at a time
No. | Item | Unit | Qty | Remarks |
1 | Oxidation tube | Piece | 6 |
|
2 | Condenser | Piece | 6 |
|
3 | Oxygen tube | Piece | 6 |
|
4 | Glass thermometer(50℃~100℃、scale division 0.2℃) | Piece | 1 |
|
5 | Bath cover | Piece | 6 |
|
6 | Silicone tube (inner diameter Φ6) (Exports to the glass through an oxygen flow meter tube -250㎜/ root) | Root | 6 |
|
7 | Silicone tube (inner diameter Φ6) (The silicone tube used for glass tube in series, water inlet and outlet, the length of each segment according to need) | Root | 3 meters |
|
8 | Cylinder connecting pipe (connecting the oxygen bottles to the instrument, with a cylinder connection) | Meter | 2 |
|